Literature Review | Electrical Impedance Tomography Monitoring in Adult ICU Patients: Recent Advances, Standardized Acquisition and Processing, Recommendations for Clinical Application, and Future Directions
01/09
2026
In the ICU, monitoring pulmonary function in mechanically ventilated patients has long been a central challenge in clinical management. Conventional monitoring methods struggle to accurately capture the dynamic changes in regional lung ventilation and perfusion, whereas electrical impedance tomography (EIT), a safe and noninvasive bedside imaging technique, is poised to deliver a revolutionary advance in respiratory care for critically ill patients by providing real-time, continuous regional assessment of ventilation and perfusion distribution. Consequently, the application of EIT is essential for evaluating pulmonary physiology and implementing personalized mechanical ventilation. Despite its promise, definitive evidence of clinical benefit remains lacking. To address this gap, a four-day expert meeting was convened in Den Bosch, the Netherlands, in the spring of 2024, to discuss optimal approaches for acquiring, processing, analyzing, and interpreting EIT data within both clinical and scientific contexts. The expert panel distilled five priority areas for recommendation from the proceedings: (1) EIT acquisition, (2) EIT signal and image processing, (3) Applications during controlled ventilation, (4) Applications during spontaneous breathing, (5) Ventilation–perfusion assessment, with a focus on adult patients in the intensive care unit (ICU). Published in 2024 in Critical Care, this expert consensus document establishes standardized operating procedures and clinical application guidelines for EIT in the ICU. This article will break down the key practical considerations and core clinical value, facilitating the rapid implementation of this technology.
01 Basic Operations: Standardized Data Collection Is the Foundation of Data Reliability
The clinical value of EIT begins with standardized data acquisition; any procedural deviation can compromise the results. The key steps must adhere to the following principles:
(1) Electrode Placement: Precise Localization Is Key
Standard position: Placed transversely between the 4th and 5th intercostal spaces (measured along the parasternal line), avoiding placement that is too high, too low, or rotated/tilted to prevent misjudgment of dorsal ventilation.
Special-case adjustments: When there are chest tubes, dressings, or skin lesions, the electrode placement may be elevated as appropriate; for patients with rib fractures, select an appropriately sized electrode pad to avoid further injury.
Optimizing electrode contact: Select the electrode belt size based on the half‑chest circumference, and ensure good electrode–skin contact by applying water, ultrasound gel, or a dedicated conductive paste. If necessary, position the reference electrode on the abdomen or shoulder, 15–20 cm away from the electrode belt.
Procedure: After installation, perform equipment calibration and signal quality checks. Only after observing a one-minute signal stabilization period may formal recording commence.
(II) Synchronous Recording and Interference Control
Synchronization requirements: It is recommended to store the data synchronously with physiological signals such as the respiratory waveform, esophageal pressure, and electrocardiogram on the same device, facilitating subsequent comprehensive analysis.
Interference avoidance: Turn off active electronic devices and prevent conductive fluids such as sweat or saline from contacting the electrodes; prohibit the use of pulsatile mattresses to minimize patient positioning changes.
Special note: Active implantable pacemakers and ICDs may introduce artifacts; device filtering settings should be adjusted accordingly. For prolonged monitoring, the electrode lead position must be marked to ensure data comparability.
(3) Key Points for Data Collection
Negative change in tidal impedance: may represent diaphragmatic motion artifacts, or may indicate clinical conditions such as pneumothorax or pleural effusion; interpretation should be guided by the overall clinical context.
End-expiratory lung impedance: suitable only for comparisons within short time intervals; it is influenced by body position, electrode contact, fluid load, and other factors, and therefore should not be used for absolute-value comparisons between patients or across time periods.
Prioritize the use of relative metrics: for example, tidal volume distribution and other relative parameters, which offer greater clinical utility than absolute values.
02 Core Technologies: Signal Processing and ROI Segmentation Techniques
(1) Signal Filtering and Lung Segmentation
Signal filtering: It is necessary to separate respiratory and cardiovascular signals. The device’s built-in frequency filters can meet routine requirements, with low‑pass filters being the most commonly used to attenuate high‑frequency cardiac signals. However, care must be taken, as an improperly chosen cutoff frequency can degrade the fine details of the respiratory signal—such as waveform shape and amplitude—thereby affecting subsequent analyses. For detailed analysis, it is recommended to export the raw data for offline processing. The goal is to remove noise and extract clean physiological signals; only after these clean signals have undergone lung‑region segmentation and ROI selection (defining specific areas of interest, e.g., for comparing dorsal versus ventral ventilation) can meaningful clinical parameters be calculated.
Lung segmentation: By setting an impedance‑change threshold (e.g., 10–15% of the maximum tidal impedance), lung tissue is distinguished from non‑pulmonary structures such as the chest wall and heart, thereby focusing on regions of functional ventilation.
(2) ROI Allocation: Each of the two approaches has its own focus.
Global ROI: This approach partitions the entire image based on its pixel matrix—e.g., by dividing it into quadrants or horizontal layers. Its advantage is that it encompasses areas with low or absent ventilation, enabling a more sensitive assessment of overall ventilatory heterogeneity; however, the anatomical regions it defines—particularly the most dependent zones—may not align with actual lung tissue due to inter‑individual variability.
Functional ROI: Segmentation is performed exclusively within the functional lung contours delineated by lung segmentation (e.g., ventrodorsal layering). This approach enables more direct physiological comparisons between dependent and non-dependent regions; however, the anatomical regions defined—such as the “dorsal” region—may not fully correspond to those identified on CT or other anatomical imaging modalities.
Key point: The choice of ROI—whether the entire lung or specific functional regions—directly influences the interpretation of ventilation distribution, such as the effects of PEEP or positional changes. Recent approaches seek to dynamically define ROIs based on each region’s relative contribution to total tidal impedance, thereby rendering regional partitioning more physiologically meaningful.
03 Clinical Applications: Comprehensive Coverage Across All Scenarios, from Mechanical Ventilation to Spontaneous Breathing
(1) Patients on mechanical ventilation: Visualized lung-protective ventilation
Individualized PEEP Titration (Core Application)
Principle: By performing a decremental PEEP trial, changes in regional compliance are monitored—loss of compliance as PEEP is reduced suggests lung collapse, while loss of compliance as PEEP is increased indicates overdistension. The intersection of the two trend lines corresponds to the “balance point” PEEP.
Key operational points: Perform the procedure in a controlled ventilation mode (volume- or pressure-controlled) to ensure stable breathing and accurately measure driving pressure; ensure the PEEP range is sufficiently broad (e.g., 6–24 cmH₂O), and expand the range further when the optimal point approaches the boundary.
2. Other applications
Intervention efficacy assessment: Real-time monitoring of regional ventilation changes in the prone and lateral positions, as well as after bronchoscopy, enables guidance on PEEP settings during the prone position and prediction of oxygenation improvement.
VILI Risk Alert: Identifies regional tidal volume heterogeneity and ventilation delay (RVD index), guiding tidal volume adjustments to reduce the risk of lung injury.
( II) Patients with Spontaneous Breathing: Weaning from Mechanical Ventilation and Optimization of Respiratory Support
1. Monitor respiratory parameters
Tidal Volume Estimation: By correlating impedance changes with lung volume, tidal volume can be noninvasively estimated; however, absolute tidal volume calculations require “point calibration” (e.g., using spirometry), and the calibration results are valid only in the short term.
Respiratory rate monitoring: Enables accurate measurement of the spontaneous respiratory rate.
Minute ventilation changes: By combining tidal volume and respiratory rate, changes in minute ventilation can be estimated.
2. PEEP titration
In assisted ventilation, EIT can be combined with regional peak inspiratory flow or dynamic transpulmonary pressure to help identify a PEEP level that simultaneously minimizes both lung collapse and overdistension. Challenges include signal stability, the need for esophageal pressure measurement, and time‑consuming analysis.
3. Risk Assessment for Weaning and Extubation
Early warning for P‑SILI (pro‑induced lung injury): identifies high-risk ventilation modes such as local overdistension and oscillatory ventilation.
Early identification of weaning failure: During the spontaneous breathing trial, a progressive decrease in end-expiratory lung impedance (indicative of atelectasis) and an increase in the ventilation heterogeneity index suggest an elevated risk of weaning failure.
4. Optimization of Respiratory Support
Noninvasive ventilation parameter optimization: Evaluate the real-world effectiveness of high-flow oxygen therapy, noninvasive ventilation, and various interface types, and tailor an individualized treatment plan.
Asynchrony Assessment: By integrating flow–pressure waveforms, it provides a visual representation of asynchrony phenomena such as trigger delay and double triggering, guiding adjustments to ventilator settings.
5. Pendulum Swing (Pendelluff) Detection
EIT is currently the only technology capable of performing bedside, real-time gas‑exchange monitoring, enabling assessment of regional time constants, resistance, and compliance; characterizing gas flow between dependent and non‑dependent lung regions; and quantifying gas distribution throughout the respiratory cycle.
(3) Assessment of ventilation/perfusion (V/Q) matching
EIT assessment of V/Q: Oxygenation depends on the match between alveolar ventilation (V) and capillary perfusion (Q). V/Q mismatch—such as well‑ventilated but poorly perfused areas, creating a “dead space,” or well‑perfused but unventilated regions, leading to “shunting”—is the underlying cause of hypoxemia in conditions like ARDS, pulmonary embolism, and pneumonia. Unlike patients requiring transport or CT imaging with ionizing radiation, EIT can be readily performed at the bedside, facilitating dynamic monitoring of disease progression and treatment response.
Procedure: Administer 10 mL of 5%–10% hypertonic saline via rapid central venous injection, while simultaneously holding breath for 8–15 seconds to eliminate cardiac pixel interference, and then calculate the pixel-level V/Q ratio.
Clinical value: Visualization of pulmonary perfusion defects and V/Q mismatch regions—e.g., in pulmonary embolism or areas of localized vasoconstriction, these appear as “ventilation without perfusion” (high V/Q zones); assessment of the efficacy of interventions such as PEEP titration, prone positioning, and inhaled nitric oxide helps elucidate the mechanisms of hypoxemia and advance personalized therapy.
Figure 3 shows an example of EIT assessment of ventilation–perfusion matching. The image data were obtained from a patient who had undergone pulmonary endarterectomy.
Key points illustrated:
A. Changes in EIT waveforms across the four lung quadrants (right ventral RV, left ventral LV, right dorsal RD, and left dorsal LD) during bolus injection of NaCl and breath-holding (38–50 seconds).
B. Bolus injection of NaCl and the distribution image of tidal ventilation prior to breath-holding.
C. Bolus‑based perfusion images: Red/yellow regions indicate changes in electrical conductivity induced by the bolus during the pulmonary perfusion phase (approximately 47 seconds). Due to a perfusion defect in the patient, these changes are observed only in the left lung.
D. EIT image after heart-rate filtering.
E. Relative EIT V/Q images within the lung region of interest: The color scale used is logarithmic, with a value of 1 indicating that ventilation and perfusion are evenly matched in that region. At present, there is no universally agreed-upon color scale for EIT perfusion or V/Q imaging.
This figure visually illustrates how EIT noninvasively and in a visual manner reveals regional pulmonary perfusion abnormalities and ventilation-perfusion mismatches.
04 Key Recommendations and the Future
(1) Key Reminders for Clinical Use
Safety: The electrolyte load from a 10 mL bolus injection of hypertonic saline is extremely low (approximately 0.5 g NaCl), and no related adverse reactions have been reported.
Standardization: Adhering to the minimum reporting standards for EIT data acquisition, processing, and interpretation (e.g., the open-source tool ALIVE) to ensure research reliability and data comparability.
Limitations: Some parameters and threshold values still require further validation through clinical studies, and their practical application should be guided by both the product’s instructions for use and clinical experience.
(II) Technical Positioning and Value
EIT does not replace conventional monitoring; rather, it provides “endoscopic‑style” regional insights, elevating respiratory mechanics monitoring from a global perspective to precise regional assessment. In optimizing ventilation strategies, guiding safe weaning, preventing ventilator‑induced lung injury (VILI), and managing hypoxemia, EIT has become an essential tool for precision‑based, personalized management in modern critical care medicine.
As standardization advances and clinical evidence accumulates, EIT is transitioning from a research tool to an indispensable bedside instrument for ICU physicians. Mastering its standardized procedures and techniques will provide stronger support for respiratory management in critically ill patients. In summary, standardization is the cornerstone of EIT’s maturation and widespread adoption, while multi‑frequency measurements, three‑dimensional imaging, and artificial intelligence constitute the key enablers propelling it from a “functional monitor” toward an intelligent bedside diagnostic platform.