Literature Highlight | A Novel Method for Individualized EIT Electrode Placement: A Clinical Study Based on Ultrasound–Diaphragm Localization


08/14

2026

PART 01

Research Background: Why is Repositioning Necessary?

 

EIT detects changes in intrathoracic electrical impedance via chest-wall electrodes, enabling real-time visualization of ventilation distribution and commonly used to guide the setting of mechanical ventilation parameters. Currently, clinical practice typically places the electrode belt between the 4th and 6th intercostal spaces; this standard originates from early consensus but does not account for individual variations in diaphragmatic position.

In fact, diaphragmatic position can vary significantly due to factors such as obesity, intra-abdominal hypertension, surgery, and respiratory failure. If the electrode belt remains fixed at its conventional location, the measurement plane may deviate from the primary pulmonary regions, leading to image distortion and information loss, which in turn can compromise clinical decision-making. Therefore, developing an electrode‑belt positioning method that accommodates individual anatomical variations holds clear clinical significance.

 

PART 02

Research Methodology

 

This study is a prospective investigation, divided into two cohorts: a healthy volunteer cohort (to establish reference standards) and a critically ill patient cohort (to validate clinical utility).

Establishment of diaphragm-based electrode belt positioning (in healthy volunteers)

This study enrolled eight healthy male volunteers and conducted measurements in the supine position. The highest point of the diaphragm, as determined by ultrasound (L0), was designated as the reference zero; cranial direction was defined as positive, and caudal direction as negative. EIT electrode belts were sequentially placed at seven transverse planes spanning from 2 cm caudal to 10 cm cranial relative to L0, with each plane monitored for 2 minutes. By comparing the amplitude of tidal impedance changes across these planes, the plane exhibiting the largest value was identified as the “optimal” electrode placement.

Comparison of the new method with the traditional method (in critically ill patients)

In a study involving 50 mechanically ventilated ICU patients—excluding those younger than 18 years, with a BMI greater than 50 kg/m², or with chest wall deformities—the authors compared a novel diaphragm‑ultrasound–guided localization method with the conventional fifth intercostal space–based approach. Under ultrasound guidance, the optimal diaphragmatic reference plane (L6cm) and the standard fifth intercostal plane were first identified, and EIT data were acquired for 2 minutes at each site. Based on whether the diaphragmatic reference plane overlapped with the traditional 4th–6th intercostal range, patients were categorized into an agreement group (n = 23) and a disagreement group (n = 27). Clinical characteristics and EIT parameters were then compared between the two groups. Using custom MATLAB software, the lung was divided into four longitudinal regions—ventral, ventrolateral, dorsolateral, and dorsal—and four cross‑quadrants, enabling systematic assessment of both global and regional ventilation metrics. This allowed for an analysis of the differences between the two electrode‑belt placement strategies in ventilation monitoring.

PART 03

Research findings

 

Determination of the Optimal Positioning Plane for Healthy Volunteers

Among the seven test planes, the global TV value was highest at L6cm (6 cm cephalad to the diaphragm’s reference point) at 4957 ± 1305 au, while the lowest TV value was observed at the L–2cm plane (1822 ± 970 au). Furthermore, the ventilation distribution across regions was more uniform in the L6cm plane, confirming that this plane optimally encompasses the lung tissue with effective ventilation.

Comparison of Global Ventilation Parameters in Critically Ill Patients

Improved global ventilation metrics: The global TV at the diaphragmatic baseline plane (3082 ± 1103 au) was significantly higher than that at the conventional five‑intercostal plane (2691 ± 1139 au), and the TV/VT ratio was also greater (7.12 ± 3.0 vs. 6.21 ± 3.0), indicating that the new method can capture more comprehensive ventilation information.

Comparison of ventilation areas and distribution: Compared with the conventional five‑intercostal‑space positioning, the ventilation area at the diaphragmatic reference plane (shaded in light blue to white) is significantly larger, while the negative‑impedance regions indicative of dead space (in purple) are markedly reduced. Moreover, this plane exhibits higher ΔZ values across the ventral (ROI1), mid‑ventral (ROI2), mid‑dorsal (ROI3), and dorsal (ROI4) regions than the conventional plane, with the most pronounced difference observed in the mid‑ventral region (250 ± 331 au). This not only underscores an expansion of the ventilated territory but also highlights an improvement in the precision of regional ventilation assessment.

Monitoring outcomes in patients with different BMIs

Patients in the discordant group had a significantly higher BMI (26 ± 3 vs. 23 ± 2 kg/m²), and their global TV and TV/VT ratios at the conventional five‑intercostal‑space plane were significantly lower than those in the concordant group. However, after adopting the new method, there was no difference in ventilation monitoring parameters between the two groups, suggesting that the new approach can correct monitoring biases in obese patients.

PART 04

Summary

 

This study provides an objective, anatomically guided, individualized approach for the placement of EIT electrode arrays and preliminarily demonstrates that it can enhance the accuracy of ventilation monitoring. As EIT becomes increasingly widespread in critical care, anesthesia, rehabilitation, and other clinical settings, integrating it with imaging modalities such as ultrasound to achieve “precise positioning” holds promise for further improving the technique’s clinical utility and reliability.