Literature Review | The Impact of Different Electrode Placement on EIT Imaging: A Prospective Observational Study
08/17
2026
PART 01
Research Background
Pulmonary complications—whether postoperative or arising during critical illness—significantly impact patient outcomes. Evidence suggests that the quality of mechanical ventilation can either exacerbate, prevent, or ameliorate such pulmonary complications; however, the precise implementation and target populations for “lung‑protective ventilation strategies” remain unclear, and one‑size‑fits‑all approaches fail to address individualized needs. Electrical impedance tomography (EIT) is a noninvasive bedside tool that monitors tidal ventilation and lung‑specific ventilation patterns, enabling the development of personalized ventilation strategies. By generating cross‑sectional images of intrathoracic impedance changes, EIT provides real‑time visualization of regional ventilation distribution and is increasingly being adopted in clinical anesthesia and intensive care.
PART 02
Research Methodology
Research subject
Sixteen mechanically ventilated patients were enrolled in the interdisciplinary ICU at the Medical School of Hanover, Germany. Exclusion criteria included chest wall deformities, active implanted devices, and skin lesions. Due to technical issues, one patient was excluded from offline analysis, leaving 15 patients who completed the study. Baseline characteristics: mean age 62.6 ± 14.8 years; male-to-female ratio 14:1; BMI 28.8 ± 4.4 kg/m² (borderline obese); SAPS II score 35 ± 10.6 (predicted mortality 10%–20%); and most patients had mild to moderate hypoxemia (PaO₂/FiO₂ ratio 255 ± 75 mmHg).
Parameter Metrics
EIT was performed using a silicone electrode belt equipped with 16 interelectrode sensors, applying a low‑amplitude alternating current of 9.1 mA at 108 kHz, with a frame rate of 20 Hz, and images were reconstructed using an improved Newton–Raphson algorithm. Ventilation settings: the patient was positioned in a 30° semi‑recumbent posture and ventilated in pressure‑controlled mode (biphasic positive airway pressure), with a tidal volume (VT) of 8.2 ± 1.8 ml/kg, calculated on the basis of predicted body weight. Monitoring parameters included all respiratory variables (respiratory rate, VT, PEEP, peak airway pressure, etc.) and hemodynamic parameters; one arterial blood gas analysis was obtained at the start of EIT. The offline ratio of tidal volume–impedance change to tidal volume (TV/VT) was calculated, and impedance distribution patterns corresponding to both automatic and user‑defined color scales were analyzed.
Electrode positioning
Starting from the 3rd–4th intercostal space just below the axilla (L1, the most cranial level), seven electrode layers (L1–L7) were established at 2‑cm intervals along the thoracic cavity, with L1 corresponding to the 3rd–4th intercostal space and L5 and below covering the paradiaphragmatic region. Two researchers jointly placed the electrode belts, ensuring that the electrode planes were perpendicular. After stabilizing each layer for 5–10 minutes, monitoring was initiated, with each layer monitored for more than 2 minutes. Respirator and EIT data were synchronously stored via the device’s bus interface.
PART 03
Research findings
Relationship between the TV/VT ratio and changes in negative impedance
TV/VT ratio and the relationship with negative impedance changes: The TV/VT ratio is highest in the L1 layer and decreases progressively in the cranio‑caudal direction, with a significant decline observed after the L5 layer (P<0.0001), representing a reduction of more than 50% compared with the L1 layer (P<0.01). Regional differences: In the dependent region (dorsal), the TV/VT ratio in lung tissue shows a significant decrease from the L5 to L7 layers (P=0.0002), whereas in the non‑dependent region (ventral), a significant decrease emerges only between the L6 and L7 layers (P<0.0001). Negative impedance changes: The area of regions exhibiting negative impedance changes increases gradually in the cranio‑caudal direction (P<0.002), with this trend being particularly pronounced at the L5 level and below.
In summary: At the L1–L4 levels, particularly between the 4th and 5th intercostal spaces, the TV/VT ratio is stable and strongly correlated with tidal volume, yielding reliable data. Below the L5 level, however, as the diaphragm enters the imaging plane, the TV/VT ratio drops sharply, the proportion of negative impedance increases markedly, and data reliability declines, making errors more likely. To obtain stable and reliable EIT monitoring data, it is recommended to position the electrode belt at the 4th–5th intercostal space (within the L1–L4 range) to minimize interference from diaphragmatic motion during measurement.
In the electrode layer, L1–L4 (more cephalad): The image is predominantly blue/white (positive impedance), primarily reflecting ventilation‑related pulmonary tissue activity. L5–L7 (more caudal): After manual color adjustment, the purple negative‑impedance regions expand markedly, indicating an increased proportion of interference signals arising from the movement of structures such as the diaphragm and mediastinal muscles.
The issue of automatic color scaling
Automatic Color Scaling: In the device’s default mode, 85% of the color range is allocated to positive impedance changes (related to ventilation), while only 15% is assigned to negative impedance changes. This results in diaphragmatic interference signals (negative impedance) in the paradiaphragmatic region being masked and appearing as black. Manual Color Scaling: After user-defined adjustments, the region of negative impedance changes (shown in purple) becomes clearly visible, enabling identification of diaphragmatic motion and assessment of mediastinal structures’ impact on ventilation monitoring.
Research indicates that, in clinical analysis, a “dual‑mode approach” is essential: first assess the overall ventilation distribution using automatic mode, then use manual mode to identify and rule out artifacts, thereby preventing the omission of critical signals.
Phase-inverted signal
In EIT monitoring, electrode placement is critical for signal accuracy. A phase-inverted signal indicates improper electrode positioning: at the paravertebral level at L5 and below, the impedance waveform (black) can be observed to be out of phase with the ventilator volume waveform (green)—with impedance decreasing during inspiration and increasing during expiration, resulting in a reversed phase relationship.
The underlying principle is that, when the electrode placement is too low, diaphragmatic motion during respiration enters the imaging plane. The resulting impedance changes exceed those associated with pulmonary ventilation, causing diaphragmatic movement to dominate the signal and thereby generating a spurious signal.
Clinical evidence indicates that a phase‑inversion signal suggests the electrode belt is positioned too close to the diaphragm; in such cases, the data acquired at that level no longer accurately reflect pulmonary ventilation and should not be used to guide adjustments of tidal volume, PEEP, or other ventilatory parameters. Upon detection of a phase‑inversion signal, the patient’s electrode belt should be promptly repositioned to the recommended 4th–5th intercostal space (L1–L4 level) to ensure the reliability of monitoring results.
PART 04
Research Conclusions
The clinical usability and reliability of EIT monitoring hinge on four key components: accurate electrode‑belt positioning, proper visualization of impedance signals, standardized data analysis, and reliable interpretation. When monitoring global parameters such as tidal volume and end‑expiratory lung volume, the electrode belt should be secured between the 4th and 5th intercostal spaces along the parasternal line, no more than 10 cm below the axillary line. The automatic color‑scale adjustment feature of EIT devices may obscure critical information; clinicians must supplement it with manual adjustments to detect negative‑impedance signals, such as diaphragmatic interference. If phase inversion or overshoot artifacts occur, first verify that the electrodes are not positioned too close to the diaphragm; after repositioning the electrodes, reassess ventilation. Data containing abnormal signals should not be used for adjusting ventilatory parameters.